BVI recalls certain ophthalmic Mini-Blades in Canada over a packaging seal defect

A clinic worker looks over sealed instrument pouches at a stockroom counter, next to a tray and a clipboard.
AI-generated illustration by AITMONDO. A healthcare worker checks stock after BVI recalled certain ophthalmic Mini-Blades over a packaging seal defect.

Health Canada lists six Mini-Blade catalogue codes in a voluntary recall that began on 2 November 2026. Healthcare facilities should check their stock and ask BVI which lots are affected. A defective packaging seal can compromise sterility.

Health Canada recall RA-82949, published 12 November 2026

Check the catalogue code and lot number, then quarantine affected stock.
AITMONDO graphic based on primary sources.

Which stock needs checking?

Health Canada published recall RA-82949 on 12 November. It lists catalogue codes 364820, 364822, 364823, BEAVER6400, BEAVER6700 and BEAVER6900. The notice gives no individual lot numbers and directs readers to the manufacturer for details. Check both the catalogue code and the lot number. See BVI product photos for three of the listed codes (PDF)

Why BVI recalled the blades

Beaver-Visitec International says some pouches in affected lots may have an open seal. The company confirmed the defect through visual inspection and dye penetration testing. A break in the sterile barrier can increase the risk of infection if the instrument is used. In its 2 November notice, BVI reported one complaint about an open seal and no related adverse health outcomes.

What healthcare facilities should do

BVI asks facilities to review their stock, quarantine affected units and stop using or distributing them. Contact the manufacturer to confirm which lots are affected and to obtain instructions for handling those units.