Medline expands recall of Hudson RCI neonatal and infant breathing circuits

U.S. healthcare facilities should review Medline’s October 7 update to the Hudson RCI neonatal and infant heated wire breathing circuit recall. It covers listed circuit item numbers with a labeled manufacturing date on or before August 31, 2026, and the Bubble CPAP Kit HUD26007KIT. Medline says lot 26GOA649 of HUD88009KIT was listed in error and is outside this recall.
Medline’s October 7 recall update, item table, images and provider instructions
Use the October 7 scope when checking inventory
Medline’s October 7 notice covers all lots of listed circuit models with a labeled manufacturing date through August 31, 2026, and adds Bubble CPAP Kit HUD26007KIT because it contains affected product. Check the device name, item number, lot number and labeled manufacturing date against Medline’s item table and recall portal. Affected items made before January 2024 carry Teleflex labels, and those made from January 2024 onward carry Medline labels. See Medline’s official label examples. The notice also corrects the September 4 update: lot 26GOA649 belongs to HUD88009KIT and is outside this recall. Some FDA records separately list individual lot codes, so facilities can check specific stock against the manufacturer’s recall portal.
Connector damage can interrupt heating
Medline reports sporadic electrical connector failures in affected circuits. The heated wire connector at the humidifier may show discoloration, melting, smoke, a burning odor or other localized heat damage. See Medline’s official product and connector damage photos in Figure 2. Medline says the humidifier’s protective system can cut power to the heated wires while respiratory gas flow continues. Inadequately heated or humidified gas and increased condensation can pose respiratory risks to infants, including airway injury or obstruction and worsening distress.
Follow Medline’s provider instructions
The October 7 notice tells healthcare providers to review inventory again under the revised criteria, including facilities that already responded or received replacement product. It provides instructions for quarantining affected stock, submitting the recall response form, and disposing of product after that response. It also addresses the exceptional situation in which use is unavoidable to prevent patient harm, monitoring during use, and actions if connector damage is observed. Medline lists QualityAssurance@medline.com and 1-800-950-0128, option 5, for quality complaints.
FDA classifies the recall as Class II
Medline began the U.S. recall on July 13, 2026, and sent further updates before the October 7 expansion. The FDA updated its public communication on September 1 to identify the action as a Class II recall. The FDA device recall database has three records for the event and listed them as open and classified when checked on November 13. Those records reflect an evolving recall; Medline’s October 7 notice supplies the current product scope and provider instructions.
Primary sources
- Medline’s October 7 recall update, item table, images and provider instructions
- FDA’s September 1 Class II recall communication
- Medline’s initial recall announcement (posted by FDA on August 6)
- FDA device recall event 99403, three enforcement records
- Reference photo by أمين (Wikimedia Commons)