Canada’s Dräger recall covers all lots of two neonatal breathing circuits

A coil of translucent breathing tubing in a sealed package rests next to a clipboard on a clinical supplies counter.
Illustration of a packaged neonatal breathing circuit in a clinical supplies area. AITMONDO (AI-generated).

Health Canada lists all lots of VentStar Resus Neo MP00310 and VentStar Resuscitaire MP00311 in recall RA-82525. Users found visible cracks in the hoses before use. The agency directs healthcare facilities to contact the manufacturer for further information.

Health Canada recall RA-82525

Health Canada lists all lots of part numbers MP00310 and MP00311 and advises contacting Dräger.
AITMONDO graphic based on primary sources.

Which products are affected?

Health Canada published the Type I recall notice on August 25, 2026, and lists August 24 as the recall date. All lots of both catalogue numbers are affected. Dräger’s August customer letter gives the device identifiers 04048675422358 for MP00310 and 04048675422341 for MP00311. See the official product photos.

What did Dräger find?

Dräger says a cracked hose can leak or detach, restricting ventilation and potentially delaying treatment. The company said it had received no reports of adverse effects on patients in the cases it cited. Its August safety notice shows the damaged hose and connector.

What should Canadian healthcare facilities do?

Health Canada directs facilities to contact the manufacturer. In its August letter, Dräger tells customers to inspect the hose according to the instructions for use and discard any circuit with a crack or detached hose. The letter describes free replacement through a local Dräger representative. Canadian facilities can ask Dräger for the response and replacement procedure that applies to their stock.